Industry Industry Industry

Understand the specific configuration of glass rotor flowmeter.

Industry

2023-07-15 10:18:45

Understand the specific configuration of glass rotor flowmeter.


Glass rotor flowmeter is used in chemical industry, petroleum, light industry, medicine, environmental protection, food and measurement, scientific research and other departments to measure the flow of single-phase non-pulsating fluid (liquid or gas).


LZBF series corrosion-resistant glass rotor flowmeter has strong corrosion resistance, and can detect the flow of acid (except hydrofluoric acid), alkali, oxidant and other corrosive gases or liquids, which is suitable for chemical industry, pharmacy, paper making, sewage treatment and other industries. The principle and structure of shape and installation dimensions The stainless steel glass rotor flowmeter is mainly composed of a conical glass tube that expands from bottom to top and a stainless steel float that moves up and down with the fluid flow.


When the fluid flows through the conical tube from bottom to top, the lift S generated by the kinetic energy of the fluid and the buoyancy A of the fluid make the float rise. When the sum of the lift S and the buoyancy A is equal to the gravity G of the float itself, the float is in balance and stable at a certain height, and the scale on the conical tube indicates the flow value of the fluid.


WB stands for with regulating valve, W stands for without regulating valve, WBF stands for anti-corrosion type with regulating valve, WF stands for anti-corrosion without regulating valve, and the reading position of the float in the flowmeter is as shown in Figure 2: the conical tube of GD -LZB/LZB-( )F flowmeter is a smooth inner wall tube, and the diameter is above DN15, and the float moves up and down through the guide rod to maintain stability; There are three guiding ribs on the inner wall of the conical tube of the flowmeter to keep the float stable.


Flowmeters with diameters below DN10 are connected by hoses and equipped with needle-shaped flow control valves; Flowmeters with diameters above DN15 are connected by flanges. Material, model, specification and technical parameters of parts contacting with measuring fluid LZB(WB) means with regulating valve, and LZB(W) means without regulating valve.


All stainless steel flanges, floats, guide rods, support plates and bolts are made of stainless steel 304(1Cr18Ni9Ti). If 316 is required as the liquid contact material, the model is GD -LZB-( )Bo, which needs to be customized. If stainless steel-lined PTFE is needed, the model is GD -LZB-( )BF (or BoF), which needs to be customized.


PTFE base, float and needle valve are made of PTFE. The supporting plates and screws are made of stainless steel LZB-( )WB and W flowmeters, which can detect the small flow of gas or liquid, and have a compact appearance. They are used for various analytical instruments, environmental protection equipment, medical equipment and other test instruments.

Specification for use of glass rotor flowmeter


Glass rotor flowmeter (hereinafter referred to as flowmeter) is one of the instruments used to measure the flow of non-turbid liquid, gas and other single-phase media. The instrument has the advantages of simple structure, convenient maintenance and use, and low price. Mainly used in chemical industry, petroleum, light industry, medicine, chemical fertilizer, chemical fiber, electric power, metallurgy, food, sugar, fuel, paper, environmental protection and scientific research departments.


The measuring part of the instrument consists of a vertically installed glass conical tube and a float in the tube. The big end of the conical tube is upward, and the float moves up and down along the axis of the conical tube with the flow. When the fluid passes through the conical tube from bottom to top, due to the action of the fluid, the upper and lower end faces of the float generate a differential pressure, and the float rises under the action of this differential pressure.


When the rising force acting on the float is equal to the resultant force of gravity, buoyancy and viscous force on the float, the float is stable at a certain height. At this time, the height of the float in the conical tube has a corresponding relationship with the flow through it. This height is a measure of the flow. The conical tube is engraved with a flow scale, and the reading of the flowmeter reads the flow indication value according to the reading position shown in Figure 1.


The structure of glass rotor flowmeter can be divided into three types according to different calibers: dial type, replaceable type and fixed type.


(1) Before installing the instrument, carefully check whether the glass tube is damaged, take out the filler to prevent the float from jumping, and check whether the float can slide freely up and down.       


(2) The flowmeter must be installed vertically on the pipeline (the big end of the conical tube is upward), and the flow direction is from bottom to top. The piping system for installing the flowmeter shall have sufficient rigidity, and the flowmeter shall not be twisted vigorously during installation.      


(3) In order to facilitate observation and maintenance, sufficient space shall be reserved during installation.       


(4) If there is backflow in the pipeline, especially when there is water hammer, in order to prevent the flowmeter from being damaged, a one-way check valve should be installed behind the downstream valve of the flowmeter. In order to facilitate inspection, maintenance, replacement of flowmeter and cleaning of pipeline, it is recommended to install bypass pipe according to Figure 5.       


(5) When the measured fluid contains larger particulate matter or is dirty, a filter should be installed upstream of the flowmeter.      


(6) When the measured fluid is pulsating flow, so that the fluctuation of the float cannot be measured, a buffer or a constant value device with appropriate size should be installed upstream of the flowmeter to eliminate the pulsation.       


(7) When used to measure gas, in order to make the instrument work stably, the pressure at the outlet end of the flowmeter should not be too low. The valve upstream of the flowmeter should be fully opened, and the valve downstream of the flowmeter should be used to adjust the flow, and this valve should not be too far away from the outlet of the flowmeter.

Link-B2B is a professional supplier listing service platform which will provide high quality and trustworthy supplier list to international buyers. If you are looking for glass rotor flowmeter suppliers, feel free to visit Link-B2B.com and send a request.


Related Post

Ship Crane Slewing Bearings: Essential Components for Maritime Operations

Slewing bearings​ play a crucial role in the opera...


Metallurgical Crane Slewing Bearings: Key Components for Heavy-Duty Operations

Slewing bearings ​are essential for metallurgical ...


A Comprehensive Overview of Slewing Bearings for Container Cranes

Slewing bearings​ are essential components in cont...


The Appeal and Versatility of Custom Sweatshirts

Custom sweatshirts offer a versatile way for indiv...


Custom Women's Shirts: Elevating Style and Personalization in Fashion

Custom women's shirts offer numerous benefits that...